Calibration - Preparation of ICE Point (Overview)
Preparing the ice point is crucial for calibrating thermometers accurately. While there might not be a specific "international guideline" for preparing the ice point, there are generally accepted best practices followed by calibration laboratories and industries worldwide. Here's how to prepare the ice point:
Equipment and Materials Needed:
- Distilled water
- Crushed or shaved ice
- Clean container (such as a Dewar flask or a thermos)
- Stirring rod
- Thermometers to be calibrated
- Thermometer with known accuracy (primary thermometer)
- Timer or stopwatch
Procedure:
Cleanliness Precautions:
- Ensure that all equipment, containers, and thermometers are clean and free of contaminants. Any impurities can affect the accuracy of the calibration process.
Prepare Distilled Water:
- Use distilled water to minimise impurities that could affect the ice point temperature.
Prepare Ice:
- Crush or shave ice into small pieces. The ice should be free of impurities and air bubbles.
Prepare the Container:
- Use a clean container such as a Dewar flask or a thermos. It should be large enough to hold the ice and water mixture with space for stirring.
Mix Ice and Water:
- Pour distilled water into the clean container, filling it to a suitable level.
- Gradually add the crushed or shaved ice to the water, stirring gently with a clean stirring rod. The goal is to achieve a slushy mixture with a uniform distribution of ice particles.
Temperature Equilibrium:
- Allow the ice-water mixture to sit for a few minutes, ensuring that the temperature stabilises. Stir occasionally to maintain uniformity.
Insert Thermometers:
- Insert the thermometer(s) to be calibrated into the ice-water mixture. Ensure that the sensing element of each thermometer is fully submerged and not touching the sides or bottom of the container.
Wait for Stabilisation:
- Allow sufficient time for the temperature readings on the thermometers to stabilise. This may take several minutes depending on the accuracy and responsiveness of the thermometers.
Primary Thermometer Comparison:
- Compare the readings of the thermometers being calibrated with the readings of the primary thermometer, which should be accurate to a known standard.
Record Temperature Readings:
- Once the temperature readings stabilise, record the readings from each thermometer. The ideal temperature for the ice point is 0°C (32°F).
Repeat if Necessary:
- If the temperature readings deviate significantly from the expected value of 0°C (32°F), repeat the process to ensure accuracy.
Documentation:
- Document the temperature readings obtained during the ice point calibration process. This documentation is essential for verifying the accuracy of thermometers and maintaining calibration records.
- Reference
- https://www.astm.org/e0563-22.html